Citation
Lee C. K., . and Low K. S., . and Hoh R., . An evaluation of an inductively coupled plasma (ICP) emission spectrometer. pp. 39-44. ISSN 0126-6128
Abstract
The performance of an Inductively Coupled Plasma (ICP) emission spectrometer was evaluated in terms o f its m ono chromator resettability ICP generator stability and detection limits. The studies showed that the analytical system exhibited similar performance to simultaneous direct reading spectrometers. The reliability o f the monochromator torch height adjustor and the stability of the source unit wereconfirmed.
Download File
Full text available from:
Official URL: http://www.pertanika2.upm.edu.my/Pertanika%20PAPER...
|
Abstract
The performance of an Inductively Coupled Plasma (ICP) emission spectrometer was evaluated in terms o f its m ono chromator resettability ICP generator stability and detection limits. The studies showed that the analytical system exhibited similar performance to simultaneous direct reading spectrometers. The reliability o f the monochromator torch height adjustor and the stability of the source unit wereconfirmed.